Research Article - (2016) Volume 2, Issue 4

Immunochemical Screening for Synthetic Cannabinoids in Workplace Drug Testing

Nadia De Giovanni1, Vincenzo Lorenzo Pascali2, Sonia Giacometti2 and Nadia Fucci1

1Institute of Public Health, Section of Legal Medicine- Catholic University – Logo F. Vito 1-00168 Rome, Italy

2Randox Laboratories, 55 Diamond Rd, Crumlin, County Antrim, BT29 4QY, UK

Corresponding Author:

Nadia Fucci
Institute of Public Health
Section of Legal Medicine Catholic
Logo F. Vito 1-00168 Rome, Italy
Tel: 003-9635507031
E-mail: nadia.fucci@unicatt.it

Received date: November 18, 2016; Accepted date: December 07, 2016; Published date: December 14, 2016

Citation: Giovanni ND, Pascali VL, Giacometti S, et al. Immunochemical Screening for Synthetic Cannabinoids in Workplace Drug Testing. J Drug Abuse. 2016, 2:4.

Visit for more related articles at Journal of Drug Abuse

Abstract

Background: synthetic cannabinoids (SC) have become a more recently abused cannabimimetic drugs; their abuse is an ongoing health issue worldwide. They are rarely detected in urine which is the most common matrix employed in workplace drug testing. Methods: this technical note refers to the investigators experience with Randox (DOA V Synthetic Cannabinoids panel Biochip Array Technology) and Concateno (Drug Screen test) immunoassays for SC, 50 authentic and anonymous urine samples were collected and analyzed from workers. Drug free urine samples to which SC were added showed the expected results in term of the declared cross reactivity’s. Results: Two urine samples obtained from workers showed positive results with Randox DOA V kit for phenyliperazines, the same urine specimens were negative with Concateno kit, which doesn’t include phenylpiperazines in the analytes panel. Additionally Concateno immunoassay found five positive samples for SC not revealed by Randox kit. Conclusion: The current paper discusses problems to be addressed before a routine investigation for SC is conducted, because immunochemical techniques are really useful only when standards, metabolites and confirmation techniques are available and well standardized.

Keywords

Synthetic cannabinoids; Immunochemical screening; Workplace drug testing

Introduction

In the past few years, the European Union (EU) and many international research centers referred to the appearance of new psychoactive substances (NPS) - including Synthetic Cannabinoids (SC) - in the illicit market [1-3]. These compounds are synthetic drugs, also called “designer drugs”, with dangerous pharmacological and toxicological effects for humans as established by Weaver et al. in 2015 [4]. The high risk is related to the presence of unknown clinical effects, including acute toxic outcome. Many clinical cases demonstrated major effects on the psychophysical performances and state of consciousness [4-6].

They are marketed to avoid current European legislation as alternatives to cannabis, often labeled “not for human consumption” however, common routes of administration include inhalation and oral ingestion [4]. Before 2008, the use of products containing SC was restricted to a small number of experimental drug users [6] and the first SC drugs detected in herbal smoking mixtures in the European market were JWH- 018 and JWH-073 [1,6]. Since 2008, many different compounds appeared in the illicit market, and their analytical identification is still difficult for the wide variability of compounds, the unknown metabolites and pharmacokinetics [7,8].

The classification of the SC, based on the chemical structures of the molecules, has been suggested by Howlett et al. [9] and Thakur et al. [10]; classical, non-classical, hybrid, aminoalkylindoles and eicosanoids.

The recent rise and widespread availability of many SC support the need for a urine screening, focused on the detection of these compounds [11,12]. Methods using LC-MS/MS or high resolution techniques for SC screening have been widely published [13]. However these techniques are not always available for the routinary analysis in all forensic laboratories, hence the employ of immunochemical screening should be helpful. Evaluation of SC use with specific drug screenings is necessary for clinical, forensic, drug treatment and workplace drug screening programs. The screening of workers employed in higher risk jobs does include drug testing analysis, with a restricted panel of the more common drugs of abuse. Workplace guidelines issued by the European Workplace Drug Testing Society (EWDTS) have defined the common drugs of abuse, their cut-off and which biological samples have to be used [14]. They don’t include the detection of the NPS, which are not under legal control in all European countries, although their increase would require a better evaluation.

Only a few forensic laboratories are equipped to identify the NPS with the immunochemical screening [15]. It is well known that immunoassay testing offers rapid separation of presumptive positive and negative specimens, prior to more costly and timeconsuming chromatographic confirmation.

The most common commercially-available immunoassays for urinary SC tests in Europe are supplied by Concateno, Randox and Neogen. This technical note refers about experience with Randox (DOA V Synthetic Cannabinoids panel- Biochip Array Technology) and Concateno (Drug Screen test) immunoassays for SC, analyzing 50 authentic and anonymous urine samples collected from workers in the year 2013. No ethical approval was necessary for the experience.

Materials and Methods

Evidence Investigator Biochip Array Technology is used to perform simultaneous detection of multiple analytes from a single patient sample. The core of the technology is the Randox Biochip; a solid state device with array of discrete testing regions containing immobilized antibodies specific to different drugs of abuse compound classes. The Randox DoA V Urine kit (Randox laboratories Limited, 55 Diamond Road, Crumlin, County Anntrim, UK) used in this paper employs a competitive chemiluminescent immunoassay, where the drug in the specimen and drug labelled with horse radish peroxidase (HRP) are in direct competition for the antibody binding sites. Increased levels of drug in a specimen will lead to reduced binding of drug labelled with HRP and thus a reduction in the chemiluminescent signal emitted. The light signal generated from each of the test regions on the biochip is detected using digital imaging technology and compared to that from a stored calibration curve. Immunochemical screening contains antibodies for mephedrone HCl (Bath Salts I assay- BSI), mescaline HCl (MESC), MDPV/MDPBP HCl (Bath Salts II-BSII), salvinorin A (SALVN), synthetic cannabinoids (SCI, SCII, SCIII and SCIV), benzylpiperazines (BZP), 1-(3-chlorophenyl) piperazine HCl (mCPP, PNPI and PNPII). Randox DOA V kit specifications provide sensitivity, limit of detection for each class of compound (Table 1a). Samples analysis has been performed as described in Randox DOA V kit insert. The Concateno Synthetic Cannabinoids Drug Screen Test (92 Milton Park, Abbingdone, Oxfordshire, OX14 4RY, UK) is intended for screening for the presence of cannabinoids in urine. It is a lateral flow immunoassay for the qualitative detection of SC metabolites in human urine, at a cut-off level of 30ng/ml. The test is based on the principle of competitive immunochemical reaction between a chemically labeled drug and the drug or drug metabolites which may be present in the urine sample for the limited antibody binding sites. Compounds producing positive results, as Concateno specification, are reported in Table 1b. 50 authentic anonymous urine specimens positive for cannabinoids (obtained by the routinely immunoassay analysis for the common drugs of abuse) were analyzed for SC using Randox and Concateno technologies. Specimens were collected over one year from people submitted to workplace drug testing and stored at -20°C until the analysis. Furthermore, four drug free urine samples were spiked with SC certified reference standards available in the Forensic Laboratory (kindly obtained by the Department of Therapeutic Research and Medicines Evaluation-Drug Abuse and Doping Unit- Istituto Superiore di Sanità) at the final concentration of 10 ng/mL to check Randox kit. In the first sample JWH-251, JWH-073 and JWH-019 were added. The second sample was spiked with JWH- 018, JWH-122, JWH-073 butanoic acid and the third with JWH- 018 pentanoic acid and JWH-081-N-5 hydroxypentyl. In the last sample JWH-073-5-hydroxyindole, JWH-250 was added. Three drug free urine samples were spiked with the same SC standards at the final concentration of 50 ng/mL to check Concateno specificity; in particular in the first sample JWH-073 and JWH-081 were added. The second sample was spiked with JWH-018 and JWH-018-N-4-hydroxypentyl. In the third sample JWH-073 butanoic acid and JWH-018 pentanoic acid were added.

COMPOUND CALIBRATION ASSAY RANGE ng/mL SENSITIVITY ng/mL LIMIT OF DETECTION ng/mL
SCI-Synthetic Cannabinoids I Assays JWH-018 0-200 1.47 3.67
SCII- Synthetic Cannabinoids II Assay JWH-018 0-200 0.87 3.69
SCIII -Synthetic Cannabinoids III Assay JWH-018 0-200 0.35 1.19
SCIV -Synthetic Cannabinoids IV Assay JWH-250 0-100 0.31 1.17
BSI- Bath Salts I Assay Mephedrone HCl 0-38 0.08 0.18
BSII- bath Salts II Assay MDPV/MDPBP HCl 0-1000 12.58 17.62
BZP- Benzylpiperazines 1-Benzylpiperazine 0-100 0.34 4.02
PNPI-Phenylpiperazines Assay 1-(3-chlorophenyl) piperazine HCl (mCPP) 0-50 0.19 1.15
PNPII -Phenylpiperazines Assay 1-(3-chlorophenyl) piperazine HCl (m CPP) 0-50 0.19 3.51
MESC- Mescaline Assay Mescaline HCl 0-250 0.65 4.07
SALVN- Salvinorin Assay Salvinorin A 0-20 0.02 0.05

Table 1a: Randox DoAV kit technical specifications.

COMPOUND SENSITIVITY ng/mL
JWH-018 pentanoic acid 30
JWH-018–N-4- hydroxypentyl 200
JWH-081–N-5- hydroxypentyl 1000
AM-2201-N-4-hydroxypentyl 1000
RCS-4-N-5-carboxypentyl 250
JWH-073 butanoic acid 15
JWH-073–N-4- hydroxybutyl 300
JWH-200–N-6- hydroxyindole 300
JWH-250–N-5- hydroxyindole 300
Lamotrigine 50

Table 1b: Compounds producing positive results with Concateno kit.

Results and Discussion

To our knowledge the most common, commercially-available immunoassays for urinary SC tests are marketed by Concateno, Randox and Neogen. In their general characteristics referred by the manufacturers, are scheduled. All the tests are specific for urine matrix, but Neogen is able to analyze blood and serum too (Table 2). Randox has a dedicated kit for SC analysis on whole blood other than urine. Concateno identifies only SC, while Randox technology can identify much more molecules. Table 3 summarizes cross reactivities of the kits for SC only; their comparison reveals that Randox can identify many more molecules compared with Concateno.

    Technology   ; Qualitative/ quantitative   Matrix   ; Assay time N. samples/ kit   Detection   Sample dilution   Sample Volume   Molecules detected
  Concateno   Lateral flow   Qualitative   Urine   6’   25   At a glance   No   N/A   JWH-018, JWH-073
    Randox       Biochip array     Semi-quantitative       Urine       30’       54       Chemiluminescence       No       25 μl   JWH-018, JWH-398,        JWH-250, Mephedrone HCl, 3’ ,4’ -Methylenedioxy-α-Pyrrolidinobutiophenone (MDPBP) HCl,                        1-Benzylpiperazine,                1-(3-Chlorophenyl) Piperazine monohydrochloride (mCPP), Mescaline HCl, Salvinorin A
  Neogen   ELISA   Qualitative Urine, blood, serum   75’   96   Absorbance   Yes   20 μl   JWH-018, JWH-073,        JWH-200, JWH-015,         JWH-019, JWH-122, AM2201, AM694

Table 2: Concateno, Randox and Neogen Kits main characteristics.

Compound Neogen Randox % CR Concateno
1-50 (ng/mL) %CR SC1 SC2 SC3 SC4 ng/mL %CR
JWH-018 0.98 100 100.0 100.0 100.0 0.7    
JWH-073-N-(4-hydroxybutyl) Metabolite 0.10 980 61.9 407.4 138.1 1.3 300 10
JWH-018 N-5-hydroxypentyl 0.13 754 227.0 415.4 227.1 0.9    
JWH-200 0.16 613 269.0 382.0 115.0 <1    
JWH-018- N-pentanoic acid 0.16 613 39.2 231.3 58.7 <1 30 100
AM2232 0.16 613            
JWH-073 0.20 490 116.1 298.5 127.5 <1    
AM1220 0.21 467 34.3 327.7 238.6 0.4    
JWH-073 N-butanoic acid 0.23 426 11.0 207.4 12.1 <1 15 200
(±) JWH-018-N-(4-hydroxypentyl) Metabolite 0.25 392 77.7 295.6 126.8 <5 200 15
AM2201 0.28 350 225.7 101.7 219.1 <1    
JWH-022 0.42 233 53.2 80.1 69.6 <1    
JWH-018 N-(5-hydroxypentyl) ββ-D glucuronide 0.49 200 18.0 308.4 65.3 0.8    
AM-2201 N-(4-hydroxypentyl) Metabolite 0.59 166 71.7 260.4 68.4 0.6 1000 3
3-(1-naphthoyl)-1H-Indole 0.64 153            
JWH-018 6-hydroxyindole 0.78 126 13.6 36.9 62.7 <1    
AM694 0.90 109 28.5 13.5 3.1 <1    
JWH-019 1.0 94 89.0 50.0 82.0 <1    
MAM2201 1.1 88            
JWH-015 1.2 83 26.3 44.5 5.1 <1    
JWH-018 4-hydroxyindole 1.6 60 30.6 3.6 10.7 <1    
JWH-122 1.9 51 71.2 2.0 9.8 <1    
JWH-018 5-hydroxyindole 2.0 50 4.9 51.8 65.5 <1    
AM-2201 6-hydroxyindole 2.0 50 6.9 72.3 54.2 <1    
JWH-007 2.9 34 16.0 17.0 2.0 <1    
JWH-398 7.5 13 20.9 <1 5.6 0.2    
WIN 55,212-3 mesylate 9.2 11 <1 11.0 8.0 0.0    
JWH-081 16 6.1 44.2 <1 <1 0.9    
JWH-210 21 4.8 51.3 <1 1.4 <1    
JWH-250- N-(5-carboxypentyl) Metabolite 51 1.9            
JWH-250- N-(4-hydroxypentyl) Metabolite 82 1.2 1.0 0.6 <1 206.0    
JWH-250 188 0.5 1.5 <1 <1 100.0    
JWH-203 205 0.5 <1 <1 <1 59.0    
RCS-4 255 0.4 61.0 <1 <1 4.1    
RCS-8 365 0.3 <1 <1 <1 0.7    
JWH 081 N-(5-hydroxypentyl)     172.3 1.5 2.5 <1 1000 3
RCS-4 N-(5-carboxypentyl)     5.5 <1 <1 <1 250 12
JWH 200 6-hydroxyindole metabolite     73.7 540.4 146.1 <1 300 10
JWH-250-N-5-Hydroxyindole             300 10
Lamotrigine             50 60

Table 3: Cross reactivities of the three different immunoassays.

A direct comparison between the different technologies is difficult due to high variability of the molecules and related metabolites. However, Randox system appears to be more sensitive than Concateno.

Drug free urine samples to which SC were had the expected results in term of the declared cross reactivity’s. Two urine samples obtained from workers showed positive results with Randox DOA V kit; the first one was positive for BSII (>30 ng/mL) and PNPII (>7.5 ng/mL), the second was positive for PNPI (>68 ng/mL) and PNPII (>68 ng/mL). The same urine specimens were negative with the Concateno kit, which doesn’t include phenylpiperazines in the analytes panel. Additionally the Concateno immunoassay found five positive samples for SC, which was not revealed by Randox kit.

The current paper discusses problems to be addressed before a routine investigation is conducted, because immunochemical techniques are only useful when standards, metabolites and confirmation techniques are available and well standardized. This is a very important aspect for the interpretation of the immunochemical results. The aim of this experimentation was also to note and underline the suggestions of EWDTS guidelines [14] that included also SC analyses in its last version.

Finally the authors want to emphasize the advice of the United Nations Office on Drugs and Crime - UNODC - [6] that promotes the collection, updating and sharing of scientific, epidemiological, forensic and toxicological information within specialists.

Conclusion

The number of abusers of SC has increased remarkably worldwide however there is an underestimation of the phenomenon. They are rarely detected in urine which is the most common matrix employed in different context as workplace drug testing.

The paper discloses problems to be underlined before the routine investigation, because immunochemical techniques are really useful when standards, metabolites and confirmation techniques are available and well standardized. Finally the aim of this experimentation was also to remember and underline the suggestions of EWDTS guidelines that included also SC analysis in the last version.

References

  1. European Monitoring Centre for Drugs and Drug addiction. www.emcdda.europa.eu
  2. ElSohly MA , Gul W , Wanas AS , Radwan MM (2014) Synthetic cannabinoids: analysis and metabolites. Life Science 97: 78-90. www.ncbi.nlm.nih.gov/pubmed/24412391
  3. Fattore L, Fratta W (2011) Beyond THC: the new generation of cannabinoid designer drugs. Front Behav. Neurosci 5: 60 www.ncbi.nlm.nih.gov/pubmed/22007163
  4. Weaver MF, Hopper JA, Gunderson EW (2015) Designer drugs 2015: assessment and management. Addiction Science Clin.Practic 10: 8 https://ascpjournal.biomedcentral.com/.../10.../s13722-015-0024-7
  5. Castaneto MS,  Gorelik DA, Desrosiers NA, Hartman RL, Pirard S, et al. (2014) Synthetic cannabinoids. Epidemiology, pharmacodynamics and clinical implication. Drug.Alcohol Depend 144: 12-41 www.ncbi.nlm.nih.gov/pubmed/25220897
  6. World Drug Report 2014. Available from: https://www.unodc.org/wdr2014/
  7. Patton AL, Seely KA, Chimalakonda KC, Tran JP, Trass M et al. (2013) Targeted metabolomic approach for assessing human synthetic cannabinoid exposure and pharmacology. Anal. Chem 85: 9390-9399. 
  8. Castaneto MS, Wohlfarth A, Desrosiers NA, Hartman RL, Gorelick DA (2015) Synthetic cannabinoids pharmacokinetics and detection methods in biological matrices. Drug. Metab. Rev 47: 124-174 www.ncbi.nlm.nih.gov/pubmed/25853390
  9. Howlett AC, Barth F, Bonner TI, Cabral G, Casellas P, et al. (2002) International Union of Pharmacology. XXVII. Classification of cannabinoid receptors. Pharmacol, Rev 54: 161?202 www.ncbi.nlm.nih.gov/pubmed/12037135
  10. Thakur GA, Nikas SP, Makriyannis A (2005) CB1 cannabinoid receptor ligands. Mini Rev. Med. Chem 7: 631?40 www.ncbi.nlm.nih.gov/pubmed/16026309
  11. Barnes AJ, Young S, Spinelli E, Martin TM, Klette KL, et al. (2014)  Evaluation of a homogenous enzyme immunoassay for the detection of synthetic cannabinoids in urine. Forensic Sci. Int 241: 27-34 www.ncbi.nlm.nih.gov/pubmed/24845968
  12. Spinelli E, Barnes AJ, Young S, Castaneto MS, Martin TM, et al. (2015) Performance characteristics of an ELISA screening assay for urinary synthetic cannabinoids. Drug Test. Analysis 7: 467-474 www.ncbi.nlm.nih.gov/pubmed/25167963
  13. Scheidweiler KB, Jarvis MJ, Huestis MA (2015) Nontargeted SWATH acquisition for identifying 47 synthetic cannabinoid metabolites in human urine by liquid chromatography-high-resolution tandem mass spectrometry. Anal. Bioanal. Chem 407: 883-97 www.ncbi.nlm.nih.gov/pubmed/25224637
  14. European Workplace Drug Testing Society - www.ewdts.org/ewdts-guidelines.html 
  15. Beck O, Rausberg L, Al-Saffar Y, Villen T, Karlsson L, et al.(2014)Detectability of new psychoactive substances “legal highs”, in CEDIA, EMIT, and KIMS immunochemical screening assays for drugs of abuse. Drug Test. Analysis 6: 492-499.